Advertisement

– There is no visible damage to device or power cables
4 Observe the instructions for use and check that all components and accessories
needed to use the product are present.
5 Check the electrical safety in accordance with the IEC 62353 or IEC 60601
standard.
12.5

Maintenance

The following table shows the preventive maintenance intervals. This table is not
related to warranty periods and no warranty is implied by the intervals listed.
Component
Clock battery
Radiometer
Spring arm joints
1) When the spring arm no longer maintains its position
The BiliLux LED panel does not need to be exchanged during the expected service
life of the device.
CAUTION
Risk of injury
The use of LED lights that are not approved by the manufacturer can affect the
safety and effectiveness of the phototherapy.
► Use only LED lights recommended by Dräger. Contact DrägerService for further
information.
12.6

Maintenance instructions

Following any maintenance procedure, perform the functional checkout procedure
on page 38.
12.6.1
Exchanging the clock battery
1 Remove the back cover from the phototherapy light.
2 Remove the old clock battery.
3 Install the new clock battery (CR2032).
4 Reinstall the back cover on the phototherapy light.
5 Set the date and time.
12.6.2
Calibrating the radiometer
Calibration must be performed by specialized service personnel. Package the
radiometer in the protective case it was shipped in, and send it to DrägerService.
Instructions for use BiliLux
Interval
Measure
Every 6 years
Exchange
Every 2 years
Calibrate
Every 1 year, or as
Tighten
1)
required
Service
Target group
Service personnel
Specialized service
personnel
Users or service
personnel
67