Safety-related information
2.9

Mechanical safety

2.9.1
Using the phototherapy light with an incubator
► Phototherapy lights placed on top of the incubator may inadvertently fall off if the
incubator is moved. Before moving the incubator, disconnect the phototherapy
light from mains power and remove it from the top of the incubator.
► Phototherapy lights placed on the top of the incubator hood may interfere with
upward travel of the height adjustment of the incubator. To prevent this
interference, always remove the phototherapy light before adjusting the
incubator height.
2.9.2
Moving the trolley
To prevent injuries and damage to the device when moving the trolley, prepare the
trolley as follows:
► Remove or secure all loose system components.
► Fold the spring arm down to a compact position.
► Lower the trolley pole to its lowest position.
► Push or pull the trolley in a straight line along the length of the trolley. Lateral or
angular movement (across the width) can result in inadvertent tip-over if the
wheels encounter any obstacle.
► Do not use the phototherapy light to move the trolley.
2.9.3
Preventing tip over of trolley
The device could tip over.
► For optimum stability, always lower the trolley to its lowest position before
moving the device.
► For optimum stability, always lock all trolley wheels.
► Do not leave the device unattended when parking on an incline.
2.10

Device failure

In case of device failure, contact DrägerService.
16
Instructions for use BiliLux