PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04
NOTE:
To minimize the effect of direct radiation on the skin probe and to obtain a more accurate baby
temperature measurement, cover the sensor with a cover. Appropriate covers include Critter Covers
Probe Cover, Care-For-Me™ Cover, or equivalent insulating foam or gel covers with a reflective surface
facing the warmer module.
d. Cover the probe with a reflective Critter Covers Disposable Probe Cover or a Care-For-Me™
Cover.
NOTE:
A higher set temperature setting does not increase rapid warming.
4. Set the Set Temperature display to the prescribed temperature.
NOTE:
Fluctuations in the Heater Power indicators or the Baby Temperature display reading can result from
air currents, obstructions of radiation to the infant, or the skin probe not being in direct contact with the
skin.
5. Verify that the Baby Temperature display reading stabilizes within 0.2°C of the Set Temperature
display.
6. To silence any Baby Temperature alarms for 10 min, press the Silence/Reset key.
NOTE:
If nursing personnel are in constant attendance and a Probe alarm occurs, use Manual Mode temporarily
if the skin probe must be disconnected until it is replaced.
7. After the cause of the alarm is corrected, any Baby Temperature, Probe, and High Temperature
alarm automatically resets. Silence the High Temperature alarm for 2 min by pressing the
Silence/Reset key.

Examination Light

1. Use the Examination Light key to turn the examination light on and off. For optimum bulb life,
switch on the examination light only as required.
WARNING:
Protect the infant eyes from strong light from the examination lamp. Failure to do so could result
in patient injury.
2. When the examination light is on, protect the infant eyes.
5 - 16