Dräger BiliLux Instructions For Use Manual page 82

Led phototherapy light
Hide thumbs Also See for BiliLux:
Table of Contents

Advertisement

Technical data
14.10.6
Enclosure port immunity
Test
Band
frequency
(MHz)
(MHz)
385
380 to 390
450
430 to 470
710
704 to 787
745
780
810
800 to 960
870
930
1720
1700 to 1990
1845
1970
2450
2400 to 2570
5240
5100 to 5800 WLAN 802.11
5500
5785
NOTE: If necessary to achieve the immunity test level, the distance between the transmitting antenna
and the ME equipment of ME system may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
1) For some services, only the up-link frequencies are included.
2) The carrier shall be modulated using a 50% duty cycle square wave signal.
3) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because, while it
does not represent actual modulation, it would be worst case.
82
1)
1)
Service
TETRA 400
GMRS 4460,
FRS 460
± 5 kHz deviation
LTE Band 13,
17
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE
band 7
a/n
2)
Modulation
Maximum
power
(W)
Pulse
1.8
2)
modulation
18 Hz
3)
FM
2
1 kHz sine
Pulse
0.2
2)
modulation
217 Hz
Pulse
2
2)
modulation
18 Hz
Pulse
2
2)
modulation
217 Hz
Pulse
2
2)
modulation
217 Hz
Pulse
0.2
2)
modulation
217 Hz
Distance
Immunity
(m)
test level
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9