Disinfecting Patient Restraints; Cleaning Patient Restraints; Disinfecting/Cleaning The Mattress; Disinfecting The In∫X - Ferno Integrated Patient Transport & Loading System iNX User Manual

Table of Contents

Advertisement

Disinfectants and cleaners containing phenolics or iodines
can cause damage. Do not use products containing these
chemicals. If bleach is used, rinse with plenty of clear
water to remove bleach residue.
Using abrasive cleaning compounds or applicators on the
iN∫X can cause damage. Do not use abrasive materials to
clean the iN∫X.
Metal buckles and slide bars can be damaged by
immersion in liquids or by washing-machine action.
Disinfect and clean only as described in this manual.
Ingress Protection Rating
The iN∫X is rated IP 56. IP stands for Ingress Protection,
and indicates the iN∫X has been tested in compliance
with dust and water infiltration standards.
Solid Particle Protection (Rating: 5): Dust-protected
Liquid Ingress Protection (Rating: 6): Protected
against water infiltration
The IP Rating system was developed by the European
Committee for Electrotechnical Standardization (CENELEC)
to show compliance of ANSI/IEC 60529, Degrees of Protection
Provided by Enclosures (IP Code).
© Ferno-Washington, Inc. / 234-3622-06 / March 2019
NOTICE
Maintenance

6.4 Disinfecting Patient Restraints

1.
Remove the patient restraints from the iN∫X.
2.
Spray the metal buckles and slide bars with a disinfectant
cleaner, following the disinfectant manufacturer's instruc-
tions for application method and contact time. Do not
immerse the buckles or slide bars in liquid.
3.
Hang the restraints to air-dry if needed. Attach only
dry restraints to the iN∫X. See "Patient Restraints" on
page 52.

6.5 Cleaning Patient Restraints

1.
Immerse the restraint webbing in a solution of mild
soap and water. Do not immerse the metal buckles
or slide bars in the solution.
Repeatedly dip the webbing in clear water to rinse. Do
2.
not immerse metal buckles or slide bars in liquid.
3.
Hang the restraint to air-dry. Attach only clean, dry
restraints to the iN∫X. See "Patient Restraints" on
page 52.
6.6 Disinfecting/Cleaning
the Mattress
1.
Remove the mattress from the iN∫X.
To disinfect: Apply disinfectant to the mattress,
2.
following the disinfectant manufacturer's instructions
for application method and contact time.
To clean: Wash the mattress with warm, soapy water
3.
and a soft cloth. Rinse the mattress with clear water.
Hang the mattress to dry, or dry it with a towel. To
attach the mattress to the iN∫X, see "Mattress" on
page 28.
6.7 Disinfecting the iN∫X
Wipe all surfaces with disinfectant. Follow the disinfectant
manufacturer's instructions for application method and
contact time. Inspect the iN∫X for damage as you disinfect it.
49