Status Indicator
Display Location
Icon
A button has been pressed longer than the maximum time
allowed.
N04
Auto-equalize problem. The iN∫X is having trouble
equalizing the legs.
N05
The battery is expecting to see charging power, but the
iN∫X is not detecting power at the ICS contacts.
N06
Both buttons are being pressed at the same time.
N07
Improper charge voltage (voltage too high).
L01
The battery is not charging. Charging power has been
detected at the ICS contacts, but the battery is not
L04
detecting a charge.
Battery over-temperature.
L06
The actuator motor has overheated. The actuators are
deactivated until the motor cools.
M01
A cable is routed from the controller to each actuator. One
or both of these cables is disconnected.
M02
With weight on the iN∫X, the legs are not moving at the
H01
proper speed, indicating:
1) Excessive weight on the iN∫X;
2) Leg obstruction;
3) Mechanical problem.
Without weight on the iN∫X, the legs are not moving at the
H02
proper speed, indicating:
1) An obstruction in the legs;
2) A mechanical problem.
Communication error.
H03
Patient-surface has been above the maximum safe transport
height longer than 30 seconds;
surface and warning triangle are red. The audible alarm
sounds a pattern of 2+3 beeps.
© Ferno-Washington, Inc. / 234-3622-06 / March 2019

9 - TROUBLESHOOTING

N04
Meaning
is disabled. The patient

9.1 Status Indicators

Status codes alert the operator to function or safety issues.
The codes are grouped into four priority levels: no priority,
low, medium, and high.
Status codes are color-coded. High-priority codes are red;
all others are yellow. When a safety issue arises, a warning
triangle is displayed next to the error code.
Refer to the chart below. Some codes indicate the need for
service. See "Parts and Service" on page 64.
Solution
Release the button.
If there is an obstruction in the legs, clear the obstruction. Use caution.
The load on the iN∫X may exceed the load capacity. Use additional help
and provide lift-assistance.
Use direct power.
Wait 30 seconds. If the error does not resolve itself in this time, there is a
problem with the ICS. Report the problem to Ferno (page 2).
Stop pressing the buttons. Press only
If the situation continues, report the issue to Ferno (page 2).
Verify the connections between the iN∫X and the ICS; verify ICS
connections to the ambulance power supply. The ICS or charger may be
plugged into an outlet providing an incorrect power voltage.
Wait 30 seconds. If the error does not resolve itself in this time, this
indicates a possible problem with the ICS, charger, iN∫X, or battery.
Report the problem to Ferno (page 2).
Allow time for the battery to cool.
Use non-powered operation to complete the transport.
Allow time for the motors to cool.
Use non-powered operation to complete the transport.
Re-connect the cable(s).
1) Remove weight from the iN∫X or provide lift-assistance.
2) Clear the obstruction from the legs. Use caution.
3) Call Ferno for service (page 2).
1) Clear the obstruction from the legs. Use caution.
2) Call Ferno for service (page 2).
Use direct power to complete the patient transport, and call Ferno for
service (page 2).
Use
to lower the iN∫X to a safe transport height.
or .