15
48
15
Countdown Timer
49
15
Countdown Timer
50
© Ferno-Washington, Inc. / 234-3622-06 / March 2019
Countdown Timer
Using the iN∫X
5.12 Direct Power: Both Legs
In this mode, both sets of legs move when
Be aware:
Auto-equalize is off. Operators must maintain control of
●
the iN∫X. Do not put the patient at a severe angle.
The weight of the patient and equipment may cause the
●
legs to move at different speeds.
Display (Figure 48):
●
White background
○
Blue legs and double-headed arrows indicates both
○
legs will move.
Red wheels indicate that the iN∫X must not be rolled.
○
If the 15-second countdown timer expires, the iN∫X
○
returns to its standard operating mode.
5.13 Direct Power: Loading-End Legs
In this mode, only the loading-end legs move when
pressed. Be aware:
Auto-equalize is off. Operators must maintain control of
●
the iN∫X. Do not put the patient at a severe angle.
●
Display (Figure 49):
White background
○
Blue loading-end legs and double-headed arrow
○
indicates only the loading-end legs will move.
○
Red wheels indicate that the iN∫X must not be rolled.
If the 15-second countdown timer expires, the iN∫X
○
returns to its standard operating mode.
5.14 Direct Power: Control-End Legs
In this mode, only the control-end legs function when
is pressed. Be aware:
●
Auto-equalize is off. Operators must maintain control of
the iN∫X. Do not put the patient at a severe angle.
Display (Figure 50):
●
White background
○
○
Blue control-end legs and double-headed arrow
indicates only the control-end legs will move.
Red wheels indicate that the iN∫X must not be rolled.
○
If the 15-second countdown timer expires, the iN∫X
○
returns to its standard operating mode.
or
is pressed.
or
is
or
43