6.19 Adjusting the Oxygen Cylinder
Holder & Cleaning the Straps
The oxygen cylinder holder can be adjusted to fit various
oxygen cylinders. The holder must be removed from the
telescoping frame to adjust the mount positions and to
remove the cylinder-fastening straps for cleaning.
REMOVING THE HOLDER
1.
Raise the backrest.
2.
Extend the telescoping frame.
3.
Remove any oxygen cylinders attached to the holder.
4.
Use a 4 mm hex wrench to remove the short screw from
each of the four corners of the oxygen cylinder mount
(Figure 65). Retain the four screws.
5.
Remove the oxygen cylinder mount and all washers (4
large, 4 small, 4 rubber).
Telescoping Frame
Properly Seated Oxygen
Cylinder Mount
58
Maintenance
Do not operate the iN∫X without the oxygen cylinder
holder. The holder must be attached to provide structural
support for the telescoping frame.
The oxygen cylinder holder must be detached from the
telescoping frame to remove the cylinder-fastening
straps for cleaning.
Oxygen Cylinder Mount Assembly
NOTICE
Short Screw (4)
Loctite® 242 - 4x
© Ferno-Washington, Inc. / 234-3622-06 / March 2019
Small Washer (4)
Large Washer (4)
Rubber Washer (4)
Hex Spacer(4)
65