Instrument Firmware and Operating System

Instrument Firmware

Instrument Firmware
The instrument firmware contains all the software required to operate all the
instrument functionality. This includes the basic instrument functionality and
all the optional functionality and modes. To see what version of the instrument
firmware is currently being used press Utility, Instrument Info, Diagnostic Info
on the instrument front panel. The Firmware Revision and Firmware Date will
be listed on the display as shown in
Figure 19-1
Diagnostic Info Display

Hardware Versus Firmware

All N5166B instruments must have firmware version B.01.80 or newer.
While it is always advised to use the latest version of the instrument firmware
that is available, there are times that an older version is required for one reason
or another. However, there are certain hardware versus firmware dependencies
that exist that need to be considered before using a version of the firmware
other than the latest. While these should all be covered in service notes, the
dependencies that existed as of the printing of this guide are shown in
Table
Table 19-1
Hardware Versus Firmware Dependencies
Assembly
A2 Vector BBG Assembly
A3 RF Assembly
A5 CPU Assembly
446
19-1.
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Figure
19-1.
a
Part Number
N5180-60280
N5180-60278
N5180-60279
0960-3295
Minimum Firmware
B.01.65
B.01.65
B.01.65
B.01.75