Measure Voltage Levels - Keysight Technologies X Series Service Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Power Supply
Troubleshooting
Figure 8-2
Power Supply Voltage LED Locations
If any of the LEDs are not on when they should be, go to the
Levels"
Supply and not the filtering on the A3 RF assembly.

Measure Voltage Levels

The troubleshooting procedure in this section is designed to be used if any of
the LEDs are not on during the
Since a power supply LED located on the A3 RF assembly could be off due to a
problem with the filtering of the power supply on that assembly, this procedure
will measure the exact DC voltage levels coming out of the A1 Power Supply
assembly where it connects to the instrument – J10 on the A3 RF assembly.
1. Position the instrument so the A3 RF assembly faces up and the A3 J10
2. Referring to silkscreen on the A3 RF assembly as well as
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
section in this chapter to verify that the problem is the A1 Power
Power Supply connector is easily accessible.
Table
8-2, measure the DC voltage levels at A3 J10.
If any DC voltages measured are not at the correct level (+/- 10%) replace
the A1 Power Supply.
"Power Supply Status
"Measure Voltage
Quick-Check".
Figure 8-3
and
173