Self Test
Running Self Test
Table 4-1
Self Test Failure Number
704
800 BB Analog Baseband Self Tests
801 - 804
900 BB RF Self Tests
901 - 905
1100 MW ALC Self Tests
1101 - 1104
1105
1106 - 1119
1200 MW Multiplier Self Tests
1201 - 1205
1300 MW Mod Filter Self Tests
1301 - 1304
1400 MW Doubler Self Tests
1401 & 1402
1500 RT Board Self Tests
1501 - 1512
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Self Tests
Go to
"Troubleshooting Self Test 704" on page 91
Replace the A2 Vector Module
Go to
"Troubleshooting Self Test 901 to 905" on page 92
Replace the A7 Micro Deck
Go to
"Troubleshooting Self Test 1105" on page 94
Replace the A7 Micro Deck
Replace the A7 Micro Deck
Replace the A7 Micro Deck
Replace the A7 Micro Deck
Replace the A7 Real-Time BBG
Resolution