Assembly Replacement

Outer Cover

Outer Cover

Tools Required

— T-20 driver

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Using the T-20 driver, loosen the two screws (1) and remove the strap
3. Using the T-20 driver, remove the two hole plugs (not shown in photo)
4. Using the T-20 driver, remove the center screws from the four rear-panel
5. Remove the four bottom feet (4) from the cover by first removing the
6. Place the signal generator on its side.
7. Place one hand on the front panel, use the other hand to lift the rear of the
The RF connector can be easily damaged. Use extreme caution to not
damage the RF connector when lifting the rear of the signal generator.

Replacement Procedure

1. Reverse the order of the removal procedures.
When sliding the outer cover back on the signal generator keep fingers
away from the back of the front panel. They can get severely pinched.
2. Recommended torque for all screws is 21 in–lbs.
336
Figure 15-1
for this procedure.
handle (2).
from the opposite side of the instrument.
feet (3).
locking keys and then lifting the tab and sliding the foot toward the tab.
signal generator and slide the outer cover (5) off to remove it from the
frame.