Assembly Replacement

A3BT1 Battery

A3BT1 Battery

This battery contains lithium. Do not incinerate or puncture this battery.
Do not install this battery backwards. To dispose of the battery in a safe
manner. Refer to
The battery is part of the power off circuitry. Removing the battery will not
cause any stored data to be lost.

Tools Required

— T-10 driver
— small flat-head screw driver

Removal Procedure

Refer to
1. Disconnect the power cord.
2. Remove the covers from the signal generator. Refer to
3. Place the signal generator flat with the A3 RF assembly facing up and the
4. Using the flat-head screw driver, remove the A3BT1 by carefully prying the

Replacement Procedure

1. Reverse the steps of the removal procedure.
To install the battery, the positive side is aligned with the positive sign on
the A3 RF assembly's battery clip.
2. Perform the post-repair procedures that pertain to this replacement
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
"Lithium Battery Disposal" on page 35
Figure 15-16
for this procedure.
page 336
and
"Inner Bottom Cover" on page
front-panel towards you.
battery out of its socket.
procedure. See
Chapter 16, "Pre and Post-Repair Procedures."
.