Keysight Technologies X Series Service Manual page 41

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Boot Up and Initialization
Potential Problems During Boot Process
4. Remove the 4 screws attaching the front frame assembly to the chassis
5. Referring to
Figure 2-2
A6A1 Front Panel Interface J6 Pinout
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
and tilt it down so that you have access to the rear of the A6A1 Front Panel
Interface board assembly.
Figure
2-2, verify the standby LED drive voltage on pin 5 of J6
on the A6A1 Front Panel Interface board measures approximately +2.1
VDC.
Is the voltage on pin 5 of J6 +2.1 VDC ±0.3 V?
a. If yes, the problem is most likely the A6A3 Power Switch Assembly.
b. If not, the problem is most likely the A6A1 Front Panel Interface
assembly or the W1 Ribbon Cable from the A3 RF Assembly to the
A6A1 Front Panel Interface assembly.
41