Service and Utility Menus
Utility Service Menu
Figure 5-14
Service Menu Remove and Install Assemblies
Install Assembly
The "Finalize" routines contained within the Install Assembly menu are meant
to perform certain required tasks after one of the assemblies listed is replaced.
These assemblies include the following:
— A2 Vector BBG Assembly (N5166B, N5172B, N5182B)
— A3 RF Assembly
— A5 CPU Assembly
— A7 Micro-Deck Assembly (N5173B, N5183B)
However, these functions are not implemented for all instruments and all
assemblies that are included on the list. The only functions that will need to be
used when an assembly is replaced are:
— Finalize RF Assembly Installation
— Finalize BBG Assembly Installation