Keysight Technologies X Series Service Manual page 120

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Service and Utility Menus
Utility Service Menu
I/Q Internal Channel Correction Calibrations
This menu is only for N5166B, N5172B, and N5182B Instruments.
This menu is intended for use in instruments that originally shipped with
firmware versions earlier than B.01.10 for them to be able to achieve
performance levels equal to instrument shipped with firmware version B.01.10
or newer. However, it can be run at any point in any instrument to optimize
performance.
120
If the App Harness is turned Off the instrument power will need to be cycled
for it to come back on.
If for some reason this software appears to be corrupt reload the instrument
firmware to reinstall the App Harness software. See
"Instrument Firmware and Operating System"
the instrument firmware.
CalTest Harness
The calibration and test harness is software in the instrument that is used to
run a variety of instrument tests and adjustments. They include:
—Self Test
—I/Q Calibration (N5166B, N5172B, N5182B)
—User Flatness Calibration
—Finalize RF Assembly Installation (N5166B, N5172B, N5182B)
—Finalize BBG Assembly Installation (N5166B, N5172B, N5182B)
—I/Q Internal Channel Correction Calibrations (N5166B, N5172B, N5182B)
—Power Verify
—ALC Modulator Bias Adjustment
If any of these are executed when the calibration and test Harness is turned
Off the following error message will be seen:
644, Firmware extension module error; Failed access to operation in
calibration extension module (number)
Under normal operation this should always be set to On. However, it is
possible to boot the instrument with this set to Off through the Main
Firmware Service Menu.
If the calibration and test harness is turned Off the instrument power will
need to be cycled for it to come back on.
If for some reason this software appears to be corrupt reload the instrument
firmware to reinstall the calibration and test harness software. See
19, "Instrument Firmware and Operating System"
reloading the instrument firmware.
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide