Functional Tests
Leveled Output Power
15.Turn the arbitrary waveform player on: Press the ARB softkey to highlight
16.Set the frequency to the first value listed in
17.Set the amplitude to 7 dBm:
18.If using a power meter, configure the power meter as follows:
19.Measure the output power.
20.Repeat steps 16 through 19 for the remaining frequencies listed in
437
On.
Press Frequency > 50 > MHz.
Press Amplitude > 7 > dBm.
a. On the power meter, press the Frequency Cal Fac button.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
17-4, and confirm that the power level at each point is within limits.
Table 17-4
Output Power With Modulation
Frequency (MHz)
50
100
250
323
512
814
1275
2025
2750
3750
5250
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Table
Output Power With Modulation
Amplitude (dBm)
7
7
7
7
7
7
7
7
7
7
7
17-4:
Table
Limits (dB)
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05
±1.05