Troubleshooting Self Test 704; Troubleshooting Self Test - Keysight Technologies X Series Service Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Self Test

Troubleshooting Self Test 704

Troubleshooting Self Test 704

Self test 704 validates the 100 MHz Reference clock signal that originates on
the A3 RF assembly and is the clock pulse for the ARB FPGA on the A2 Vector
assembly. If the 100 MHz clock is inadequate in amplitude or frequency the
cause likely originates from the A3 RF assembly.
If self test 704 fails, see the
"100 MHz Reference Signal"
section in
Chapter 12,
"Baseband Generators"
to verify the level of the 100 MHz reference signal.
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
91