Amplitude Modulation; Pulse Modulation; Self Tests; A7 Micro-Deck Related Adjustments - Keysight Technologies X Series Service Manual

Signal generators
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Micro-Deck Assembly
A7 Micro-Deck Assembly Troubleshooting

Amplitude Modulation

If there are any amplitude modulation issues, see
Issues (Option UNT)"

Pulse Modulation

If there are any pulse modulation issues, see
UNW)"

Self Tests

There is a series of Self Tests related directly to the A7 Micro-Deck assembly.
They are the 1100, 1200, 1300, and 1400 series tests. See
Test,"
Table 4-1
the failure.
Before replacing an A7 Micro-Deck assembly for a Self Test failure be sure to
run the
be sure to verify the power supply voltages by seeing the
Status Quick-Check"

A7 Micro-Deck Related Adjustments

1. MW ABUS Calibration
2. MW Multiplier Prelevel Calibration
3. MW Linearity Calibration
4. MW Low Band Absolute Power Calibration
5. MW High Band Absolute Power Calibration
234
in
Chapter 6, "Troubleshooting"
in
Chapter 6, "Troubleshooting"
for a description of each of the tests. If there are Self Test failures, see
for instructions on how to proceed with determining a resolution for
"A7 Micro-Deck Related Adjustments"
section in
This adjustment is used to calibrate the internal diagnostics measuring
ADC.
This adjustment is used to determine the optimal power out of the A7A2
20 GHz Frequency Multiplier. If the power is too low, it will impact the
ability of the system to deliver enough output power. If the power is too
high, it will increase the harmonic content of the output.
This adjustment is used to linearize the ALC detector for the best power
accuracy.
This adjustment measures the absolute power at the mid-point of the ALC
vernier and generates a cross calibration file for the ALC calibration and
attenuator calibration. This adjustment also calibrates the first two
attenuator steps (bypass, 0 dB and 5 dB), covering the frequency range of
9 kHz to 50 MHz.