Functional Tests
Leveled Output Power

N5166B, N5172B, and N5182B Alternative Test Procedure

If a USB power sensor is not available, use a power meter to measure the
output power of the signal generator.
Test Setup
1. Zero and calibrate the power sensor to the power meter:
2. Connect the equipment as shown below:
3. Preset the signal generator: Press Preset.
4. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
5. Turn modulation off:
6. Continue the procedure from
7. Continue the procedure from
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Press Mod On/Off so that the Mod On/Off LED turns off.
the output power measurements of the N5172B/82B.
"Without Modulation" on page
"With Modulation" on page 436
435.
to complete