Remove The Outer And Inner Signal Generator Covers; Remove The Outer Cover; Remove The Inner Top Cover - Keysight Technologies E8251-60384 Installation Notes

Psg signal generators add option 1e1 attenuator, for option 513 and 520/532 and 540/550 and 567
Table of Contents

Advertisement

Remove the Outer and Inner Signal Generator Covers

Remove the Outer Cover

Refer to
Figure
1.
1. Disconnect the power cord.
2. Using a T-20 driver, loosen the two screws (1) and remove the strap handle (2).
3. Using a T-15 driver, remove the center screws (3) on the four rear-panel feet (4).
4. Remove the four bottom feet (5) from the cover by lifting the tab and sliding the foot toward
the tab.
5. Place the signal generator on its side.
6. Tilt the signal generator forward and slide the outer cover (6) back to remove it from the frame.

Remove the Inner Top Cover

Refer to
Figure
2.
1. Using a T-10 driver, remove the twelve screws (1) from the inner-top cover (2).
2. Remove the inner-top cover.