Commissioning The Compact Controller For Inductive Con- Ductivity; Measured Variables - ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual

With measuring /control unit
Table of Contents

Advertisement

Commissioning the Compact Controller for Inductive Conductivity
13
Commissioning the Compact Controller for Inductive Con‐
ductivity
13.1

Measured variables

Switching between measured varia‐
bles
Measured variable: Inductive conduc‐
[ConI]
tivity
Measured variable: Resistance
66
User qualification: trained user, see
n
fication' on page 15
WARNING!
Sensor run in period
This can result in dangerous incorrect metering
Consider the run in period of the sensor during
commissioning:
The controller can process the following measured variables:
Inductive conductivity
n
Resistance
n
TDS value
n
[SAL]
Salinity
n
Use the
key to switch between the controller's measured varia‐
[ConI] , [RES] , [TDS] and [SAL] in the continuous display.
bles
Depending on the measured variable set, the settings of variables
are changed or the variables are hidden completely in the
[INPUT Ø TCOMP] menu and in the [LIMIT] menu.
Symbol displayed in the controller's display:
Units of measurement: μS/cm, mS/cm , S/cm. The measuring
range is automatically detected and switched by the controller
Physical variable: specific electrical conductivity (K). Only this
measured variable and the correction variable
emitted at the mA output, independently of the measured variable
set on the controller. The setting of the measured variable on the
controller only affects the layout of the display and not the output at
the mA output.
[RES]
Symbol displayed in the controller's display:
Units of measurement: MΩcm, kΩcm, Ωcm. The measuring range
is automatically detected and switched by the controller
Physical variable: specific electrical resistance.
Calculating the specific resistance: ρ (T
Correct measuring and metering is only pos‐
sible if the sensor is working perfectly.
It is imperative that you adhere to the run in
periods of the sensor.
Calculate the run in period when planning com‐
missioning.
It may take a whole working day to run-in the
sensor.
Observe the sensor operating instructions.
[ConI]
[RES]
[TDS]
Ä Chapter 2.5 'User quali‐
[ConI]
'Temperature' are
[RES]
) = 1/K (T
)
ref
ref