ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual page 104

With measuring /control unit
Table of Contents

Advertisement

Calibrating the controller diaLog DACb
104
CAL O2
Determine reference value
07.51 ppm
Change with <OK>
Fig. 71: Reference value
6.
Use OK and the arrow keys to enter the reference value
ð The display appears as follows:
CAL O2
Reference value
07.51 ppm
Range:
0.00 ... 24.00 ppm
Fig. 72: Reference value
7.
Continue with CAL.
8.
The display appears as follows if calibration is successful:
CAL O2
Calibration successful
Slope
Before calibration
After calibration
A
Zero point
Cancel with <ESC>
Fig. 73: Successful calibration
9.
Use CAL to confirm
n
Press ESC to cancel
n
10.
The display appears as follows if calibration is unsuccessful:
Continue with <CAL>
End with <CAL>
A2859
A2858
A2854