ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual page 80

With measuring /control unit
Table of Contents

Advertisement

Calibrating the controller diaLog DACb
Tab. 8: Valid calibration values
Evaluation
Very good
Good
Acceptable
80
NOTICE!
Poor sensor operation and fluctuating pH values
during the process
The calibration method with an external sample
has a number of disadvantages compared with the
buffer solution calibration method. If the pH value
fluctuates significantly during the process, then the
pH value may change by a variable amount in the
period between sampling, sample measurement
and entry of the pH value into the controller. This
could mean that the pH value entered into the con‐
troller does not correspond to the actual pH value
in the process. Consequently the result is a linear
displacement of the pH value across the entire
measuring range.
If the pH sensor no longer reacts to changes in the
pH value and only gives out a constant uniform mV
signal, this cannot be detected using the calibra‐
tion method with an external sample. With the cali‐
bration method with two buffers (e.g. pH 7 and pH
4), this becomes apparent if the pH sensor does
not detect any changes in the pH value.
The calibration method with an external sample
should only be used with installations where there
is poor access to the pH sensor and the identical
or very uniform pH values are used in the process.
In addition the pH sensor should be regularly serv‐
iced or replaced.
Correct sensor operation
Zero point
-30 mV ... +30 mV
-45 mV ... +45 mV
-60 mV ... +60 mV
Continuous display ➨
Correct measuring, control and metering is
only possible if the sensor is working perfectly
Refer to the sensor's operating instructions
Slope
56 mV/pH ... 60 mV/pH
56 mV/pH ... 61 mV/pH
55 mV/pH ... 62 mV/pH

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dulcotrol dwca dacb

Table of Contents