Calibrating The Fluoride Sensor - ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual

With measuring /control unit
Table of Contents

Advertisement

Calibrating the controller diaLog DACb
15.1.3

Calibrating the Fluoride Sensor

15.1.3.1
Selection of the calibration process for fluoride
88
Continuous display ➨
CAL ORP
Offset
Last calibration
CAL setup
Calibration process
continue with <CAL>
Fig. 50: Data input, ORP sensor calibration (CAL)
1.
Continue with
CAL ORP
Offset
Accept with <CAL>
Fig. 51: Adjusting the [Offset]
2.
Press
and use the four arrow keys to adjust the mV value
of the buffer you are using
3.
Press
4.
Transfer the result of the calibration into the controller
memory by pressing
ð The controller operates with the calibration results.
To calibrate the controller there are two available calibration pro‐
cesses:
1 point
n
2 point
n
0.0 mV
21.05.2013 14:59:56
Data input offset
0.1 mV
A1032
A1033