Calibrating the controller diaLog DACb

15
Calibrating the controller diaLog DACb
15.1

Calibration

70
User qualification: instructed user,
n
cation' on page 15
Measuring channel settings
This descriptions of [Channel 1] apply correspond‐
ingly to the settings in all other measuring chan‐
nels. The procedure for entering the settings for
the channel in question is identical, however the
parameters to be set may differ. Your attention is
drawn to the differences, which are also described.
Display tolerances
Display tolerances between the sensor and/or
measuring device and controller must be calibrated
with sensors and/or with output signals from meas‐
uring devices that do not require calibration or
where calibration is performed in the sensor/meas‐
uring device. The relevant information for this is
contained in the respective operating instructions
for the sensor or measuring device.
Cancelling the calibration process with ESC
ESC can be used to cancel an ongoing calibration
process at any stage. The controller then con‐
tinues to operate with the last calibration result
detected as valid.
Continuous display ➨ Menu ➨
or
Continuous display ➨
.
Calibration
Please select channel
Channel 1
Channel 2
Fig. 31: Please select the channel.
Ä Chapter 2.5 'User qualifi‐
[Calibration] ➨ .
or
Chlorine
pH [mV]