16.1.2
Calibration of the temperature coefficient
CAL TEMPCOEFF
TCOEFF=0.00%/°C
CAL=OLD
Fig. 80: Calibration of the temperature coefficient
I.
If the temperature change is greater than 2°C,
the message changes to
II. If the temperature is within the specified range,
the message changes to
Calibrating the compact controller for inductive conductivity
Conductivity sensors with temperature element
You can only calibrate the temperature coefficient
with conductivity sensors with a temperature ele‐
ment, because it is impossible to calculate the tem‐
perature coefficient without measuring the tem‐
perature.
Temperature change
It is recommended that the temperature is changed
by no more than 0.5 °C per minute, or with a tem‐
perature change of e.g. 10 °C you will need to wait
for a minimum of 20 minutes before calibration.
[WAIT]
[STABLE ?]
CAL
CELLCONST
TCOEFF
CAL TEMPCOEFF
CAL_T1= 22.0 °C
CAL=CONTINUE
CAL TEMPCOEFF
CHANGE TEMP
T2 > 32.0 °C
or
T2 < 12.0 °C
I.
CAL TEMPCOEFF
WAIT
T2 > 32.0 °C
or
T2 < 12.0 °C
II.
CAL TEMPCOEFF
STABLE ?
CAL_T2= 43.0 °C
CAL=CONTINUE
III.
CAL TEMPCOEFF
ACCEPT ?
CAL_T2= 43.0 °C
CAL=CONTINUE
III. When a stable final temperature has been
reached, the message changes to
Calibration can now be terminated manually.
A1866
[ACCEPT?] .