Technical data and operating parameters
20.1
Flow and operational test of flow control
Tab. 12: Switching point of the flow control
Sample water
Designation
1
Potable water
2 and H
Rinsing water
Industrial water
Process water
4 and H
Waste water,
clear
5 and H
Waste water
with solid par‐
ticle fraction,
turbid
7
Waste water
with fluoride
and pH < 7
128
Set the ranges of the reed switch on the flow control DGMa and/or
DLG to Minimum and Maximum, switching the DACb from Opera‐
tion to Pause when doing so
Check:
the wiring to the controller is correct,
n
the switching function of the flow sensor,
n
the correct configuration of the controller,
n
the alarm relay (e.g. horn).
n
Bypass fitting
Nominal cross
section
mm
DGMa
4
DLG III
6
DLG III
6
DLG III
6
DLG III
6
Permissible flow
Control switches
off,
l/h
at l/h
20... 60
< 20
40... 65
< 20
40... 65
< 20
40... 65
< 20
40... 65
< 20