ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual page 82

With measuring /control unit
Table of Contents

Advertisement

Calibrating the controller diaLog DACb
15.1.1.4
Calibration of the pH Sensor (CAL) by
Tab. 9: Valid calibration values
Evaluation
Good
Acceptable
82
[Data Input]
Data input
With the [Data input] calibration method, the known
data of the sensor is entered in the controller. Cali‐
bration by data input is only as accurate and reli‐
able as the method with which the data was deter‐
mined.
The sensor data must have been determined
recently. The more up-to-date the sensor data, the
more reliable is this calibration method.
Correct sensor operation
Measuring and control behaviour of the controller
during calibration
During calibration the actuating outputs are deacti‐
vated. Exception: a basic load or a manual control
variable has been set. This remains active. The
measured value output [standard signal output mA]
is frozen, corresponding to its settings in the mA
output menu.
When calibration/testing has been completed suc‐
cessfully, all of the error checks relating to the
measured value are restarted. The controller saves
all the determined data for zero point and slope
upon a successful calibration.
Zero point
-30 mV ... +30 mV
-60 mV ... -30 mV
or +30 mV ... +60 mV
Continuous display ➨
Correct measuring and metering is only pos‐
sible if the sensor is working perfectly.
Refer to the sensor's operating instructions.
Slope
-55 mV/pH ... -62 mV/pH
- 40 mV/pH ... - 65 mV/pH