ProMinent DULCOTROL DWCa Assembly And Operating Instructions Manual page 102

With measuring /control unit
Table of Contents

Advertisement

Calibrating the controller diaLog DACb
15.1.5.2.2
Calibrating the slope using a reference measurement in water
102
7.
Wait for the stability of the measured values.
8.
The display appears as follows if calibration is successful:
CAL O2
Calibration successful
Slope
Before calibration
After calibration
A
Zero point
Cancel with <ESC>
Fig. 67: Successful calibration
9.
Use CAL to confirm.
n
Press ESC to cancel.
n
10.
The display appears as follows if calibration is unsuccessful:
CAL O2
Calibration not possible
Slope
Before calibration
After calibration
A
Zero point
Cancel with <ESC>
Fig. 68: Unsuccessful calibration
11.
Use CAL to end.
n
Press ESC to cancel.
n
12.
Check the sensor and installation once again and repeat the
calibration procedure.
1.
Press the CAL key in the continuous display.
2.
Use OK to select the measuring channel.
ð The display appears as follows:
End with <CAL>
---
End with <CAL>
A2854
i
A2855