Calibration - Jadever JCAII Manual

Hide thumbs Also See for JCAII:
Table of Contents

Advertisement

6) Press key
7)Press key

7-5 Calibration

= remote platform
= main scale
Note:
Please calibrate the main scale first.
The calibration steps of main scale and remote platform are the same. Press key
switch the calibration for the main scale or remote platform before the zero point calibration.
You can only do the calibration for the remote platform after you have set the max. capacity
and division of the remote platform.
Here take JCAII-6K as an example.
1. Press key
is displayed in the Unit Weight window.
2. With no load on the weighing pan, press key
till
is displayed in the Unit Weight window. Input the first calibration weight value by the
numeric key (6K model, 1/3 of full load is 2kg). And then put the corresponding weights on.
3... Press key
put the corresponding weights on.
4. Press key
the corresponding weights on.
to save the settings and return to weighing mode.
to change its unit: kg, lb, g.
and hold while turning on the scale. Do not release
and wait till
and wait till
to start the zero point calibration and wait
is displayed in the Unit Weight window. And then
is displayed in the Unit Weight window. And then put
V03
to
till
JCAII 7