Chapter 2 Set Up; Models; Calibration - Jadever JWE-3K Service Manual

Table of Contents

Advertisement

JADEVER

2-1 Models

On the A/D PCB, jumpers TPI and TP2 are used to determine the capacity model, as shown
below: (Do know that in different capacity model, load cell used may be different.)
Capacity
3kg
6kg
15kg
30kg

2-2 Calibration

On the A/D PCB, jumper CAL and jumper TC are used to determine the ON/OFF of the
calibration function and TC (temperature compensation), respectively, as shown below:
Jumper
CAL
TC
As per the right of proprietary intellectual property, may not be reproduced in any form without written permission.

CHAPTER 2 SET UP

Open-circuited
Open-circuited
Note!: Metrological controls and/or regulations
The regulations and/or metrological controls governing to conduct
calibration vary from country to country.
Please do follow the local metrological controls and/or regulations
imposed by the local authority.
Calibration/TC ON
open
shorted
TP1
Shorted
Shorted
Calibration/TC OFF
JCE/JWE
Service Manual - V05
TP2
Shorted
Shorted
Open-circuited
Open-circuited
Shorted
open
8