6. Calibration

1.For best results, calibrate the scale at regular intervals. Temperature changes, geographic
gravity variations, altitude changes and abuse are few reasons why a scale may need
recalibration.
2. Here we take JWE-3K as an example.

6-1 Single Point Calibration

MODE
1)Press and hold key
while powering on the scale. The screen displays "CAL", then
release the keys.
TARE
2)Press key
to enter the zero point calibration mode.
MODE
3)Press key
to select calibration weight value. Options are1/3 of full load, 2/3 of full
load and full load. E.g., options for JWE-3K are 1, 2 and 3(kg).
4)Put the corresponding weights on the weighing pan, then the calibration procedure
starts.
5) The calibration procedure is completed with a beep and"
" flashing. Now, remove
all the weights.
ZERO+
MODE
6)Press key
to return to weighing mode, key
to advance to next setting
menu.

6-2 Linear Calibration

TARE
1)Press and hold key
while powering on the scale. The screen displays
,
then release the keys.
JWE
7