4. Adjustments and Settings

4-1 Model setting

1) Turn on the power while pressing key SMPL.
2) Key in 1130 with the numeric keys.
3) Again press key SMPL to confirm.
4) Press numeric key
5) Press numeric key
6) Press key SMPL, then press key ZERO to go back to weighing mode.

4-2 Single Point Calibration

1) Press and hold key SMPL while powering on the scale.
2) Input 11 with numeric keys.
3) Again press key SMPL to enter the zero point calibration modes.
4) Wait till "
" flashes, press numeric key
1/3 of full load, 2/3 of full load and full load. E.g., options for JCL-15K are 5, 10 and 15(kg).Put
the corresponding weights on the weighing pan.
5) Press key
to confirm.
·
6) The calibration procedure is completed with a "
remove all the weights.
7) Press key SMPL to save, and then press key ZERO to return to the weighing mode.

4-3 Linear Calibration

1) Press and hold key TARE while powering on the scale.
2) Again press key TARE to enter zero point calibration mode, with "
weight screen.
12
Manual
JADEVER SCALE CO., LTD
1
to set maximum capacity. Available options are:3/6/15/30(kg).
2
to select version. Available options are: TW/CN.
1
to select calibration weight .Options are
" flashing on the weight screen. Now,
JCL Service
" flashing on the