Use the numeric key to set the format directly.
SET11
12)
Set the filtering level in which the stable indication turns on .The higher the setting, the
slower stabilization time.
SET12
13)
stable :
Taring or zeroing can be not action without stable indication after pressing the key
TARE
ZERO
or
always:Taring or zeroing can be action without stable indication after pressing the key
TARE
ZERO
or
auto:
Press key
SET13
14)
On: display the last weight when powering on again
Off: do not display the last weight when powering on again
SET14
15)
Qty.ck: start amount checking
Wt.ck: start weighing checking
SET15
16)
G N/SET
press key
displaying OK.

6. Single and Multi-point Calibration

Note: Before calibration, please set the capacity first. The unit used in calibration is the one
that has been set before. During the calibration procedure, LONG press
return to normal weighing mode without saving.
Here we take 3kg/0.2g as an example
TARE/
1. Press and hold
Do not release it till the window displays
2. With no load on the weighing pan,
TARE/
press
3. Wait till the window displays the first calibration value.
Note: The first calibration value is default. If the value
has been changed, the default value is 1/3 of full load.
filtering setting: 1,2,3,4
Precondition for
zeroing or taring setting: stable, auto, always
TARE
ZERO
or
weight memory parameter: on,off
the kinds of checking: Qty.ck,Wt.ck
parameter initialization
M+/SAVE
then key
while powering on.
to start zero point calibration.
, but it works only when get the stable indication.
to start initialization. It is finished when
CAL"
ZERO/ESC
to
8