= There will be a warning sound when the material quantity is less than the lower limit, and
the weight of the material is more than 20 divisions.
= No sound alarm.
8) Press numeric key
upper & lower limit of quantity checking.
.
= Previously set quantity checking values are not retained when the unit is turned on.
.
= Previously set quantity checking values are retained when the unit is turned on.
9) Press numeric key
Prt.Pr, Prt.Co and Prt.St. If Prt.Co is chosen as the print
mode, PC will be automatically selected as the
external devices.
.
= manual print
.
= continuous print
.
= Stable printing (the weight of the weighted articles should be more than 9 divisions.
Weighted articles should be removed and the scale goes back to zero before print out the
next record.)
MC/CK
10) Press key
Options are PC,AX,TP,Godex ,SH,
ET
,TDP and ZEBRA.
MR
11) Press Key
to switch On or OFF RTC
function.
12) When setting is completed, press key

6. Calibration Mode

1) If good linearity of the load cell was not guaranteed, linear calibration is required. After the
linear calibration is made, there is no need to make single point calibration.
2) If there is limit in weights, single point calibration is needed .it would be more accurate to
calibrate the scale with 1/3 or more of the full load.
3) Here we take 75kg weighing capacity as an example.
8
to set whether to save the
9
to set print modes .Options are
to select external devices.
ZERO
JWI-700C
to save and return to weighing mode.
9