Acai; Simple Recipe Function - Jadever JCAII Manual

Hide thumbs Also See for JCAII:
Table of Contents

Advertisement

3. Remove the samples and put on the articles, the scale begins to count
Note: Symbol
is less than 10 pieces.

8-3-3 ACAI

Automatic Counting Accuracy Improvement (ACAI) results in a more accurate count by
increasing the reference weight without the need to count additional parts. A higher
reference weight is important when there is a risk of inconsistent piece weights or if the
reference weight is close to the minimum. ACAI uses an initial averaging unit weight to
count additional weight that is placed on the scale. After a few seconds, the scale gives a
beep as a new higher reference weight is used to recalculate the averaging unit weight. The
process can be repeated as long as the additional weight is less than the less than previous
reference weight. Once the limit is exceeded ACAI is turned off.
Note:
If you set the average parameter (P4) to OFF, it means that the scale will do the ACAI in all
situations.
If you set the average parameter to 15, the scale will not work if there is an object shortage or
more than ±15% of the unit weight.
Whether it works or not, depends on the object's tolerant of weight.

8-3-4 Simple Recipe Function

1. Hold the
advance parameter. Press key
to activate or disable this function.
2. Press
to save the setting and return to normal weighing mode.
3. Activate the simple recipe function, and press
then the scale will tare automatically when keeping the item. Repeat the same operation until
get all material needed for a recipe.
4. Press MR to check all accumulation items.
5. Clear the weighing pan and press TARE to continue a second recipe weighing.
appears pointing at "
for 3 seconds under the normal weighing mode, then press
" in the Unit Weight window when sample quantity
or
to select parameter A-10, and press
after put the item on the weighing pan,
V03
to set the
or
JCAII 11