Adjusting Flow Quantity; Information About The Service Display Of Atmos S 61 Servant From Software Version V 1.2 - Atmos S 61 Servant Operating Instructions Manual

Ent workstation
Hide thumbs Also See for S 61 Servant:
Table of Contents

Advertisement

4.0 Operation

4.4.7 Adjusting flow quantity

Prior to stimulation, turn the set screw
2 adjusting possibilities:
• High: flow quantity for ear irrigation
• Low: flow quantity for stimulation of the vestibular organ
Prior to ear irrigation, turn the set screw to the position "High",
push it into the handle, and then adjust it by screwing it in further.
For fine adjustment of the flow, turn the adjusting screw
out.
) Periodically check the water flow for the stimulation of the
vestibular organ and readjust if necessary.

4.4.8 Information about the service display of ATMOS S 61 Servant from software version V 1.2

The module ATMOS S 61 Servant ENT workstation has a service display on the front panel.
The display consists of a red and a yellow LED.
After switching on the module, the electronics go through a test cycle. The following indications are possible:
Immediately after start-up, the LED test starts.
Start-up
Subsequently, a functional and aggregate test follows. For this purpose, six flashing signals light up (time-ratio light up/pause:
500 ms / 500 ms).
Operating time pressure pump
Operating time suction pump
Operating time suction pump 2
Operating time hose pump
Operating time drainage pump
Annual maintenance
Battery voltage
Possible maintenance cases:
Operating time pressure pump
Operating time suction pump
Operating time suction pump 2
Operating time hose pump
Operating time drainage pump
Annual maintenance
Battery voltage
If one of the above maintenance cases occurs, please contact your local ATMOS service partner.
4
to the position "Low".
in or
11
red LED
yellow LED
Lights up
Lights up
First light up
Second light up
Third light up
Fourth light up
Fifth light up
Sixth light up
Seventh light up
First light up
Second light up
Third light up
Fourth light up
Fifth light up
Sixth light up
Seventh light up
11
4
red LED
yellow LED
Beep
Remarks
Both LEDs light up simultaneously
Yes
during start-up (duration: 500 ms).
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
No
o.k.
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Critical operating time
Yes
Annual maintenance is due
Yes
Low battery voltage