4.7

Therapy progress

26
The ATMOS® C 051 Thorax has 2 graphical diagrams to
simplify the analysis of the air-flow and actual vacuum
progress.
Selection menu
By selecting the
button you reach the graphical
diagram modus. By pressing the buttons you can
select the modus of you choice e.g. long / short time.
4.7.1

Short time

The graphical diagram starts by selecting the menu.
In this modus the real measurements (flow, vacuum)
from the last 30 seconds can be shown. Therefore you
can visualise cough tests and other proceedings.
By pressing the
button the diagram can be frozen
to enable a graphical interpretation. When you
press the button
again the short time diagram is
restarted.
By pressing the
button you will return to the main
menu.
4.7.2

Long time

In the longtime modus the complete therapy progress
can be visualised.
• The scaling can be switched between time, flow and
vacuum.
• You can reach the different scalings by pressing the
,
or
button.
• The scale can be increased or decreased by
pressing the
or
• Time scaling:
- The endpoint on the right side of the graphic is
always the actual point of time.
- The scaling can be selected in 7 steps, between
the display of the past 60 minutes and the last
12 days.
- A vertical line shows when the therapy was
interrupted.
• Flow scaling:
- The scaling can be selected between 0 – 0.55 l/
min and 0 – 5.5 l/min in 4 steps.
• Vacuum scaling:
- The scaling can be selected between 0 - 110
mbar (= cm H
O) and 0 - 22 mbar (= cm H
2
steps.
button.
O) in 3
2