Prior to first use please peruse the chapter 2.0 “For your safety”, in order to be prepared for any possible The quality management system applied at ATMOS has dangerous situations. been certified according to international standard ISO 13485. The basic principles are: ...
1.0 Introduction Intended use Function Name: ATMOS C 361 The ATMOS C 361 is a mains-operated surgical suction ® ® unit, centering around a silent diaphragm-type pump Main functions: which generates a vacuum inside the secretion canister, Suction of secretions, rinsing fluids and temporarily allowing secretions to be withdrawn and collected. Using collection of body fluids.
1.0 Introduction Explanation of symbols Attention, refer to operating instructions ! Serial number Order number Follow operating instructions (blue) Protection against penetration of IPX 1 Fuse damaging humidity (drop water) Potential equalization Professional disposal Applied part Type BF Eurasian conformity Alternating current GOST Certificate (Russia) Unit of protection class II Unit Off Unit On...
Liquids must not be allowed to enter the device. Should liquids have penetrated into the device, it must be ATMOS is not liable for personal injury and damage to inspected by an authorized service technician before property if being used again.
3.0 Setting up Always set the equipment up on a secure, level surface. Bild 1. Operating elements On/Off switch with pilot lamp Vaccuumgauge Vacuum controller Bild 2. Vacuum connection: Direct-Docking-System The vacuum connection between the pump and the secretion canister is created automatically as soon as the DDS canister is positioned correctly.
3.0 Setting up Connection area in unit base Connect mains cable Use only mains cables with angled inlet connector for non-heating appliances. Check that the voltage and frequency ratings of the power line are similar to those indicated on the device. Fig.
4.0 Operation 4.1 Insert / remove DDS bacterial filter / oversuction stop Please wear gloves while changing the filter! Fig. 5. Using the DDS splash protector Fig. 6. Attach / remove DDS secretion canister lid With the DDS secretion canister on a firm surface, position the lid horizontally on top (the lid may not be twisted!) Press down lightly onto the secretion canister using both hands until limit is reached.
4.0 Operation Attach DDS secretion canister handle Insert the DDS secretion canister handle into the grooves of the lid with the snap-in hooks open. Fig. 9. Close / open DDS secretion canister handle To close, secure the snap-in hooks under the edge of the secretion canister, and then press the clips downwards until they lock into place.
4.0 Operation Insert DDS hose adapter Press the required DDS hose adapter with 6 or 10 mm diameter into the ”patient” hole of the DDS secretion canister lid twisting slightly to ensure a tight fit. Twist slightly in the same manner when removing. Fig. 12. Connect hose Fig.
The DDS bacterial filter / oversuction stop is disposable. Before each use, check that the DDS bacteria filter / oversuction stop is clean and dry. Replace the DDS bacterial filter with a new filter if it is dicoloured, contaminated or oversucked. Use only original ATMOS bacterial filters! Never operate the unit without the DDS bacterial filter / oversuction stop! Replace the DDS bacterial filter with every cleaning respectively desinfection of the DDS canister systems.
5.0 Options Trolley with standard rail A trolley with standard rail, which can also be used with disposable systems if necessary, is available for mobile use. Always position the trolley on a flat, sturdy surface. Fig. 16. 5.1.1 Securing the unit It is only possible to ensure safe operation as a mobile suction unit by using the special trolley available for use with the unit! ...
5.0 Options 5.2 Use of suction unit with disposable systems Optionally the suction unit may also be used as a tabletop unit with a disposable system it can be attached to a standard rail. This requires the standard rail adapter with vacuum connector.
6.0 Cleaning General information on cleaning Remove the DDS bacterial filter from the DDS secretion and disinfection canister handle and dispose of it. All other parts, except the bacterial filter, must also be thoroughly rinsed under running water. You may add a For disinfection, you may use all surface and instrument detergent, if you wish. disinfectants listed in chapters 6.4 / 6.5.
The unit does not require any further maintenance. At least every 24 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS recommends an inspection according to the manufacturer‘s specifications. Repairs The following may require repairs from the manufacturer or an authorized service partner. Prior to sending in the device, please contact your service partner by phone.
Send the product to ATMOS or to your dealer. 8.0 Trouble-shooting The ATMOS C 361 was subjected to a thorough quality control before shipment. If there is, nevertheless, some malfunction, ® you possibly might solve this problem yourselve if you observe the following instructions:...
DDS-splash protection................. 340.0056.0 DDS-hose adapter set, Ø 6 + 10 mm ............340.0057.0 9.2.2 For ATMOS C 361 with trolley ® Trolley with standard rail ................320.0070.1 DDS-standard rail adapter with vacuum connection for the use of disposable systems at the unit ..........340.0059.0 Grad.
10.0 Technical specifications Air flow rate of pump 36 ± 4 l/min Max. vacuum at sea level -91kPa ( -910 mbar or 682,5 mmHg)*@ NN Vacuum readout -1...0 bar ± 16 mbar(class 1.6) Additional air regulation Mechanical regulating valve, ball vacuum regulator Secretion canister 1.5 l or 3 l canisters made of polysulphone Suction hose ø...
11.2 Reprocessing In case secretion was sucked into the device it may not be operated until it is repaired by the ATMOS service. Handling of the suction device determines to a large extent its reliability and safety. The hygiene measures described in the previous chapters are necessary measures for the protection of patients and users, and to maintain functional reliability.
The ATMOS C 401 and ATMOS C 361 are intended for use in the electromagnetic environment specified below. ® ® The customer or user of the ATMOS C 401 and ATMOS C 361 should ensure that they are used in such an ® ® environment. Emissions Test Compliance Electromagnetic Environment - Guidance...
Page 24
The ATMOS C 401 and ATMOS C 361 are intended for use in the electromagnetic environment specified below. ® ® The customer or user of the ATMOS C 401 und ATMOS C 361 should ensure that they are used in such an ® ® environment. IEC 60601- Electromagnetic Environment Immunity Test...
Page 25
C 361 ® ® The ATMOS C 401 and ATMOS C 361 are intended for use in electromagnetic environment in which ® ® ratiated disturbances are controlled. The customer or user of the ATMOS C 401 and ATMOS C 361 can help ®...