ATMOS
C 051 Thorax
®
English
317.0000.B
2016-03 Index: 06

Advertisement

Table of Contents
loading

Summary of Contents for Atmos Thorax C 051

  • Page 1 Operating instructions ATMOS C 051 Thorax ® English 317.0000.B 2016-03 Index: 06...
  • Page 2 Further information, accessories, consumables and spare parts are available from: ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Straße 16 79853 Lenzkirch/Germany Tel. + 49 7653 689-0 Fax: + 49 7653 689-190 + 49 7653 689-292 (Service Center) atmos@atmosmed.de www.atmosmed.de...
  • Page 3: Table Of Contents

    6.0 Functions ............................... 31 6.1 Hose rinsing ....................................31 7.0 Accessories, consumables and spare parts ..................32-34 7.1 Attachment of the universal bracket ............................32 7.2 Attaching/removing the ATMOS C 051 Thorax to/from the universal bracket ............33 ® 7.3 Connecting of shoulder belt ..............................34 8.0 Cleaning and care............................
  • Page 4: Introduction

    These operating instructions must always be kept available near the device. Care and safety inspections in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS ® C 051 Thorax and are therefore a must besides regular cleaning.
  • Page 5: Intended Use

    1.2 Intended use 1.2.1 Intended use ATMOS C 051 Thorax ® Name: ATMOS C 051 Thorax ® Main functions: Digital device for mobile thoracic drainage Med. indications/ application: Recovery of the neutral vacuum in the intra pleural cavity after the occurrence of a pneumothorax or a pleural effusion by drainage of air and secretion.
  • Page 6 The secretion canister and the drainage hose are sterile and disposable and can be applied in the sterile OT area Contraindications: No application of the canister with other low vacuum devices than ATMOS C 051 Thorax. ® No separate application of the canister as gravity drainage.
  • Page 7: Function

    1.3 Function The ATMOS C 051 Thorax is an exceptionally handy, mobile, digital thoracic drainage suction device. ® The device is operated via an electrical, maintenance-free membrane pump. During operation the pump creates a vacuum within the suction hose and the secretion canister due to which secretion and air can be drained by the hose system.
  • Page 8: Transport And Storage

    1.4 Transport and storage The transport of the device may be effected only in a dispatch carton upholstered and offering sufficient  protection. Please document and report damages in transit immediately.  For complaints or return deliveries, please use the enclosed form QD 434. The form QD 434 can also be downloaded from: www.atmosmed.de The unit must be allowed to stand for up to six hours at room temperature prior to starting up for the first ...
  • Page 9: Explanation Of Symbols

    Check check correct fi t Pictures contained in this manual Warning,  Important information special diligent notice! Symbols of ATMOS C 051 Thorax & Accessories ® Application part The CE sign shows that Multiple use of components marked type BF...
  • Page 10: Safety Advice

    Prior to first use compare the normal voltage of the device (see  indication label on the backside of the device) with the local mains voltage. NOTICE ATMOS cannot guarantee perfect functioning neither will it be liable for damage to people or property  any non-original ATMOS parts are used, ...
  • Page 11: Notice

    The removal of the canister from the device during the therapy may  only be performed by trained professionals which act in conformity with internal regulations. The ATMOS C 051 Thorax is a medical device which is subject to ® ...
  • Page 12: Caution

    2.2 Caution CAUTION Danger of injury A misplaced drainage system and a misplaced patient hose could  hinder the drainage of air and liquids. A complete blocking of the system during the drainage of liquids and air could cause a rise in pressure and thus lead to a pressure pneumothorax.
  • Page 13 Don´t take a shower / bath with the device.  WARNING Danger of explosion due to unobserved ambient conditions The ATMOS C 051 Thorax may only be operated in rooms used for ®  medical purposes, but not in areas subject to explosion hazards or in oxygen rich environments.
  • Page 14: Setting Up And Starting Up

    3.0 Setting up and starting up 3.1 Scope of delivery The ATMOS C 051 Thorax was subjected to thorough testing prior to packaging and shipment. ® On receipt of the goods please check the package for any possible damage and compare the contents for completeness ( see bill of delivery).
  • Page 15: Device Overview

    ATMOS C 051 Thorax Cover cap USB-B (no function for user) Connection for USB fl ashdrive (therapy data-transfer) Only use the USB-connection for the transfer of therapy data. A software update may only be performed by ATMOS or an authorized service person.
  • Page 16: Starting Up

    C 051 Thorax the battery must be fully charged. ® Only the enclosed battery recharger (GTM 21097-5012) from ATMOS may be used. Advice on the handling of the battery can be found in chapter 9.3. Correct handling of the rechargeable batteries prolongs their maximum service life.
  • Page 17: Secretion Canister

    3.3.2 Secretion canister Always use the original ATMOS single use secretion canister  Vacuum connection system  The vacuum connection between device and canister is set up immediately after connection! For hygienic reasons we advise you to exchange hoses and canister ...
  • Page 18: Exchanging The Secretion Canister

    3.3.2.4 Exchanging the secretion canister 1. Press the blue operating element ( device overview) on top of the device to release the canister. 2. Hold the secretion canister slightly angled and pull upwards to remove 3. Place the secretion canister safely on an even surface. 4.
  • Page 19: Operating

    Target vacuum Display of the preadjusted target vacuum to which the pump adjusts Day/Night Mode The ATMOS C 051 Thorax has a day/night ® Key lock activated mode, the device adjusts automatically to the environmental lighting. Flow displayed as bubbles Each additional coloured bubble represents an additional flow.
  • Page 20: Buttons And Display Symbols

    4.2 Buttons and display symbols 4.2.1 Buttons Figure Function Decrease target vacuum / in the menu: decrease selected value Increase target vacuum / in the menu: increase selected value Graphic diagram of the therapy Open the user settings Save entry Confirm information Back / Exit menu Warning / suppress the warning...
  • Page 21: Switch On

    Therapy on hold Therapy in progress Upcoming warning suppressed Safety check required 4.3 Switch on 1. To switch on the ATMOS C 051 Thorax please touch the ® sensor above of the symbol for 2 seconds. 2. The welcome monitor appears with the software version number in the bottom right corner.
  • Page 22: Leakage Test

    Alternatively the thorax catheter can be clamped near to the patient. Do not clamp the ATMOS hose system. If the leakage test is error-free, the message „Leakage test OK“ appears. Now you can remove the plug from the hose.
  • Page 23: Function

     By pressing the button the pump will be stopped.  The ATMOS C 051 Thorax has a vacuum adjustment. This means, that the integrated pump only runs, ®  if the actual vacuum doesn't correspond to the target vacuum. On the other hand the pumps performance depends on the difference between the actual vacuum and the target vacuum.
  • Page 24: Release Of Keylock

    4.6 Keylock The ATMOS C 051 Thorax has an automatic keylock. ® 1. Automatic activation of the key lock If the settings are not changed for a defined time, the keylock will be activated automatically (default factory setting 5 minutes, individually adjustable in the user settings). This will prevent unintentional operation.
  • Page 25: Therapy Progress

    4.7 Therapy progress The ATMOS C 051 Thorax has 2 graphical diagrams to ® analyse the development of Air-Flow and actual vacuum. Selection menu By selecting the button you reach the graphical diagram modus. By pressing the buttons you can select the modus of your choice e.g.
  • Page 26: Transfer Of Therapy Data

    4.7.3 Transfer of therapy data You may transfer the therapy data to a USB flashdrive. The therapy data is saved as a PDF- and Excel-file. If you continue the therapy after the data transfer, the data is recorded. The transmitted data is not deleted. If you are starting a new therapy, the previous data will be overwritten.
  • Page 27: Reading Out The Therapy Data

    Following information can be seen in the report: beginning and end of recording, flow at beginning and end of recording file name und device ID graphical diagram of therapy data Therapy Report ATMOS® C 051 Thorax File name: ATMOS_C_051_THORAX_ID0C1C_20110212_063107.pdf Device ID: 0C1C...
  • Page 28: Switch Off

    4.8 Switch off To switch off the ATMOS C 051 Thorax stop the therapy and touch the ®  sensor for 2 seconds. The enacting monitor appears and the device shuts down.  4.9 User settings For accessing the user menu please press the control panel.
  • Page 29 When the device is started the standard-vacuum is automatically preadjusted. Standard - vacuum The standard-vacuum can be adjusted with Period time of hose rinsing can be Period time of hose rinsing adjusted with Key lock activation time can be Keylock activation time adjusted with The keytone can be activated or Keytone...
  • Page 30: Warning Messages

    > Contact the ATMOS service sucked into the pump. The measurement of an excessively > Check for blockages high vacuum results in the display of >...
  • Page 31: Functions

    If the target vacuum is adjusted over-50 mbar a notice appears ,,High target vacuum is set“ 6.0 Functions 6.1 Hose rinsing The ATMOS C 051 Thorax has an automatic hose rinsing function which works periodically. ®  The rinsing process transports secretion located in the secretion hose into the secretion canister.
  • Page 32: Accessories, Consumables And Spare Parts

    7.0 Accessories, consumables and spare parts 7.1 Attachment of the universal bracket (Accessories) Turning the universal bracket: The universal bracket can be attached – horizontally and vertically - to plates (e.g. table boards), pipes and stands with a diameter up to 40 mm. Pull the fi xing pin from the fi xation at the bottom side of the ...
  • Page 33: Attaching/Removing The Atmos ® C 051 Thorax To/From The Universal Bracket

     Ensure that the thread at the bottom of the vacuum unit is positioned directly above the fi xing screw of the stand holder. Turn the fi xing screw clockwise to fi x the ATMOS C 051 Thorax. ® ...
  • Page 34 Take a carabiner  and attach it to one of the clips on the drainage unit.  Attach the second carabiner  to the second clip on the drainage.  Now the ATMOS C 051 Thorax can be worn over your shoulder. ® Attaching the drainage unit to a patient’s bed: To attach the device to a patient’s bed exchange the carabiner ...
  • Page 35: Cleaning And Care

    8.0 Cleaning and care 8.1 General information on cleaning and disinfection Prior to cleaning Medical devices such as the ATMOS C 051 Thorax must always be ready for use and must function reliability. ® Therefore prior to every application: necessary The way the device is used determines its reliability and safety.
  • Page 36: Cleaning Of The Device Surface

    Attention! The device should never be autoclaved, rinsed under running water or immersed into any liquids! 8.3 Recommended disinfectants Disinfectant Ingredients (in 100 g) Manufacturer ATMOS Green & Clean SK Alkyldimethylbenzylammoniumchloride <1 g Metasys, Rum (Austria) (application solution) Dialkyldimehtylammoniumchloride <1 g Alkyldimethylethylbenzylammoniumchloride <1 g...
  • Page 37: Hygiene Plan

    8.4 Hygiene plan WHAT WHEN NOTES Manual wipe cleaning Drainage unit Manual wipe disinfection Single use product -> not Canister suitable for reprocessing exchange after use Single use product -> not Hose system suitable for reprocessing exchange after use R= Removal, C= Cleaning, D= Disinfection, S= Sterilisation...
  • Page 38: Maintenance And Service

    To carry out these measures the person must have the necessary test devices and original spare parts. ATMOS recommends: work should be carried out by an authorized ATMOS service partner. This ensures that repairs and testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
  • Page 39: Handling Of Rechargeable Batteries

    On principle batteries are depleted after 2.5 years. New batteries should be fully charged prior to first use. As ATMOS has no influence on the handling of the device, batteries are excluded from the general guarantee. A function guarantee of 6 months is effective.
  • Page 40: Eliminating Errors

    10.0 Eliminating errors Description Possible causes Measures Connect the battery recharger to the device to recharge the battery, Device cannot be switched on Battery completely empty Note that the progress can be checked in the upper left hand corner of the display Fuse defect Control fuses Battery does not recharge.
  • Page 41: Technical Data

    O, kPa and flow in l/min Data memory Internal memory for therapy data: 2.5MB up to 12 days recording possible Canister ATMOS disposable canister, transparent, pressure relief valce, ® graduation Max. volume of 800 ml, connection to the device with "direct docking system".
  • Page 42 IP 33 Applicable standards EN 60601-1 EN 60601-1-2 EN ISO 10079 Classification acc. to appendix IX EC Directive 93/42/EEC CE marking CE 0124 GMDN-Code 36787 UMDNS-Code 10-218, suction device thoracic 317.0000.0 ATMOS C 051 Thorax ® Issue of Technical Specifications: 15.12.2015...
  • Page 43: Disposal

    In order to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer or send it directly to ATMOS MedizinTechnik GmbH & Co. KG for a professional disposal.
  • Page 44: Notes On Emc

    The ATMOS C 051 Thorax is intended for use in the electromagnetic environment specifi ed below. The customer or ® user of the ATMOS C 051 Thorax should ensure that it is used in such an environment. ® Electromagnetic Environment...
  • Page 45 13.2 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS C 051 Thorax is intended for use in the electromagnetic environment specified below. The customer or ® user of the ATMOS C 051 Thorax should ensure that it is used in such an environment. ® Electromagnetic Envi-...
  • Page 46 The ATMOS C 051 Thorax is intended for use in the electromagnetic environment specified below. The custo- ® mer or user of the ATMOS C 051 Thorax should ensure that it is used in such an environment. ® IEC 60601-...
  • Page 47 To determine the electromagnetic environment in regard to stationary transmitters, a study of the location must be considered. If the measured field strength at the location where the ATMOS C 051 Thorax is used exceeds the above compliance level, the ATMOS C 051 Thorax must be ®...
  • Page 48 ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch / Germany Phone: +49 7653 689-0 atmos@atmosmed.de www.atmosmed.com...