Therapy Process; Short Time Display - Atmos C 051 Thorax Operating Instructions Manual

Table of Contents

Advertisement

4.8

Therapy process

34
The ATMOS C 051 Thorax offers 2 graphs to simplify
the analysis of the flow and actual vacuum process.
Selection menu
By selecting the buttons
graphical diagram modus. By pressing the buttons
you can select the modus of you choice e.g. long time
display / short time.
4.8.1

Short time display:

The graphical diagram starts by selecting the menu.
In this modus the real measurements (flow, vacuum)
from the last 30 seconds can be shown. Therefore you
can visualize cough tests and other proceedings.
By pressing the button
to enable a graphical interpretation. When you press
the symbols
again the short time display time
diagram is restarted.
By pressing the
button you will return to the main
menu.
) Set the period duration of the hose rinsing to > 5
minutes if you want to use the short time display
for real-time display of the flow, such as for cough
tests, to detect blocked catheters, etc.
you enter the menu for
the diagram can be frozen