Connecting the hose system

18
8. Release the 2 Luer-Lock connections by a counter-clockwise
1
4
rotation to separate the secretion canister from the hose
system. Pay attention as secretion could be found in the
connection space.
9. Remove the blue cover cap
connection with it (secretion channel ).
6
10. Remove the black cover lid  and cover the pop-off valve
with it .
11. Remove the yellow cover cap
the drainage unit with it (bacterial filter ).
12. Correctly dispose of the secretion canister.
Reinserting the secretion canister
13. Insert the new secretion canister (see chapter „3.3.2.3 Insert
secretion canister" on page 17).
14. Open the clamp at the thoracic catheter.
and cover the upper Luer-Lock
and cover the connection to
Measuring and rinsing channel
Hydrophobic bacterial filter
Luer-Lock 4 mm
Secretion channel
Luer-Lock 6 mm