4.8 Therapy process

4.8.1 Short time display

) Set the period duration of the hose rinsing to > 5 minutes if you want to use the short time
display for real-time display of the flow, such as for cough tests, to detect blocked catheters,
etc.

4.8.2 Long time display

The ATMOS S 201 Thorax offers 2 graphs to simplify the analysis
of the flow and actual vacuum process.
Selection menu
By selecting the buttons
diagram modus. By pressing the buttons you can select the
modus of your choice e.g. long time display / short time.
The graphical diagram starts by selecting the menu. In this
modus the real measurements (flow, vacuum) from the last 30
seconds can be shown. Therefore you can visualise cough tests
and other proceedings.
By pressing the button
a graphical interpretation. When you press the
the short time display time diagram is restarted.
By pressing the
button you will return to the main menu.
In the long time display the complete therapy process can be
visualised.
• The scaling can be switched between time, flow and vacuum.
• You can reach the different scalings by pressing the
buttons.
• In the scaling display the scale can be increased or decreased
by pressing the
or
Time scaling:
• The endpoint on the right side of the graphic is always the
actual point of time.
• The scaling can be selected in 7 steps, between the display of
the past 60 minutes and the last 12 days.
• A vertical line shows when the therapy was interrupted.
Flow scaling:
• The scaling can be selected between 0 – 100 ml/min and 0
– 15 l/min in 4 steps.
Vacuum scaling:
• The scaling can be selected between
0 - 100 mbar (= cmH2O
0 - 20 mbar (= cmH2O
you enter the menu for graphical
the diagram can be frozen to enable
buttons.
O) and
2
O) in 3 steps.
2
button again
,
or
35