Atmos C 451 Operating Instructions Manual

Atmos C 451 Operating Instructions Manual

Surgical suction unit

Advertisement

English
340.0339.B
340.0338.B
340.0334.B
340.0333.B
340.0300.B
2012-09 Index: 15
ATMOS C 451
Surgical Suction Unit
MedizinTechnik

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the C 451 and is the answer not in the manual?

Questions and answers

Summary of Contents for Atmos C 451

  • Page 1 MedizinTechnik English ATMOS C 451 Surgical Suction Unit 340.0339.B 340.0338.B 340.0334.B 340.0333.B 340.0300.B 2012-09 Index: 15...
  • Page 2: Table Of Contents

    4.13 Test DDS bacterial filter / oversuction stop ..12 11.0 Checking / Reprocessing / Disposal ....24 4.14 Checking wear and tear parts ......12 11.1 Checking ATMOS suction devices ......24 11.2 Reprocessing ............24 11.3 Disposal ..............24 12.0 Notes on EMC ..........25-27 General Standard Terms and Conditions ATMOS MedizinTechnik GmbH &...
  • Page 3: Introduction

    Care and safety inspections in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS C 451 and are therefore a must besides regular cleaning. Repair work and safety inspections may be carried out only by expert personnel authorised by ATMOS.
  • Page 4: Explanation Of Pictures And Symbols

    Pictures contained in this manual Warning, especial diligent  Observe operating instructions Important information notice Keys on the control panel / symbols at the ATMOS C 451 Application part type BF Potential equalization Alternating current Serial number Order number Protection against penetration of...
  • Page 5: Scope Of Supply

    1.5 Scope of supply ● Prior to dispatch, this ATMOS device was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note). In addition to the basic device, the scope of delivery comprises the following parts:...
  • Page 6: For Your Safety

    The ATMOS C 451 fully complies with the electromagnetic  The ATMOS C 451 may be operated only in rooms used immunity requirements of standard IEC 60601-1-2 / EN for medical purposes, but not in areas (zones M and G) 60601-1-2 "Electromagnetic compatibility - Medical Elec-...
  • Page 7: Setting Up

    3.0 Setting up  Always set the equipment up on a secure, level surface. Fig. 1. Operating elements  On/Off switch with pilot lamp  Vacuumgauge  Vacuum controller    Fig. 2. Vacuum connection: Direct-Docking-System  The vacuum connection between the pump and the collection jar is created automatically as soon as the DDS jar is positioned correctly.
  • Page 8: Connection Area In Unit Base

    3.0 Setting up Connection area in unit base Connect mains cable  Use only mains cables with angled inlet connector for non-heating appliances!  Check that the voltage and frequency ratings of the power line are similar to those indicated on the device. Fig.
  • Page 9: Operating

    4.0 Operation 4.1 Insert / remove DDS bacterial filter/ oversuction stop  Use gloves after having operated the unit! Fig. 8. Using the DDS splash protector Fig. 9. Attach / remove DDS collection jar lid  With the DDS collection jar on a firm surface, position the lid horizontally on top (the lid may not be twisted!) ...
  • Page 10: Attach Dds Collection Jar Handle

    4.0 Operation Attach DDS collection jar handle    Fix the container handle  into the guiding of the lid with opened snap-in hook .   Fix the snap-in hook  below the container edge and press at the upper part  towards the centre of the container until the handle snaps in.
  • Page 11: Insert Dds Hose Adapter

    4.0 Operation Insert DDS hose adapter  Press the required DDS hose adapter with 6 or 10 mm diameter into the ”Patient” hole of the DDS collection jar lid twisting slightly to ensure a tight fi t.  Twist slightly in the same manner when removing. Fig.
  • Page 12: Suction

    -0.3 bar when the vacuum controller is in the ”max.” position and the suction hose is open. The filter must then be replaced.  Replace the DDS bacterial filter at least once a day. Use only original ATMOS bacterial filters!  Never operate the unit without the DDS bacterial filter / oversuction stop! 4.13 Checking wear and tear parts...
  • Page 13: Options

    Fig. 25. System trolley  A system trolley, which can be used with 2 ATMOS system jars on a changeover adapter or with disposable systems if necessary, is available for use in the OT. ...
  • Page 14: Securing The Unit

    5.0 Options 5.2.1 Securing the unit  It is only possible to ensure safe operation as a mobile suction unit by using the special system trolley available for use with the unit!  The suction unit is placed on the system trolley so that its feet lock into place in the holes of the unit carrier and it can be firmly attached to the unit carrier from underneath by means of a knurled screw.
  • Page 15: Dds Changeover Docking Station

    5.0 Options DDS changeover docking station Max load on station: 15 kg! Max load on station: 15 kg! Damage on device at higher weight! Damage on device at higher weight!  The DDS changeover docking station for 2 collection jars with changeover switch and direct docking may be used on the system trolley.
  • Page 16: Tray In Foot Of System Trolley

    5.0 Options Tray in foot of system trolley  A tray (available as an optional extra) may be inserted in the foot of the trolley. Fig. 34. Potential equalisation  A potential equalization connector may be installed on the system trolley (in the installation groove) as an optional extra.
  • Page 17: Cleaning

    6.0 Cleaning 6.1 General information on cleaning and disinfection Prior to cleaning Medical devices must always offer a maximum in safety and function. We therefore recommend: Prior to each application: necessary Cleaning and sterilizing hoses and  The measures described herein regarding cleaning collection jar and disinfection do not replace the regulations valid for operating the device!
  • Page 18: Cleaning And Sterilizing The Unit Surface

    6.0 Cleaning Cleaning and sterilizing the unit surface  Always disconnect the device from the power line, before cleaning and disinfecting the surface.  Wipe the surface clean with a cloth soaked in a cleaning solution or disinfectant. Liquids must not enter the device. All of the cleaning solutions and disinfectants listed below can be used.
  • Page 19: Maintenance

    REF 008.0081.0 Fig 39. 8.0 Trouble shooting The ATMOS C 451 was subjected to a thorough quality control before shipment. If there is, nevertheless, some malfunction, you possibly might solve this problem yourselves if you observe the following instructions: Problem...
  • Page 20: Accessories, Consumables And Spare Parts

    340.0062.0 Change: after 60 x autoclaving Standard rail set for system trolley 340.0081.0 (instead of DDS-docking system for 2 system jars) System trolley, basic version for ATMOS C 451 340.0070.0   Tray for system trolley (134 °C) 340.0084.0 Change: after 60 x autoclaving 340.0082.0...
  • Page 21 Accessories, consumables and spare parts Accessories for ATMOS C 451 Receptal® package 2 x 1.5 l 444.0027.0 Receptal® package 1 x 2 l 444.0030.0 Receptal® package 2 x 2 l 444.0028.0 Receptal® package 1 x 3 l 444.0031.0 444.0029.0 Receptal® package 2 x 3 l Receptal ®...
  • Page 22 Accessories, consumables and spare parts Description   DDS-jar handle, PSU ......... 340.0055.0  DDS-bacterial filter / oversuction stop, hydrophobic, disposable part ..... 340.0054.0  DDS-hose adaptor set, Ø 6 + 10 mm ..340.0057.0    DDS-jar lid with gaskets ......340.0053.0 ...
  • Page 23: Technical Specifications

    10.0 Technical specifications Air flow rate of pump 45 l/min +3/-5 l/min Max. vacuum -91kPa ( -910 mbar oder 682,5 mmHg)*@ NN Vacuum readout -1...0 bar ± 16 mbar (Klasse 1,6) ø 63 mm Additional air regulation mechanical regulating valve Collection jar 1.5 l, 3 l or 5 l jars DDS or Receptal ®...
  • Page 24: Checking / Reprocessing / Disposal

    A regular check of the condensate-controller on the rear side is necessary. Pull out the plastic plug at check the colour at the end of the hose. In case of discolouration/deposits a maintenance measure must be performed by a certified ATMOS service partner! 11.2 Reprocessing In case secretion was sucked into the device it may not be operated until it is repaired by the ATMOS service. 11.3 Disposal ● The ATMOS C 451 is not comprised of any hazardous materials.
  • Page 25: Notes On Emc

    12.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS C 451 is intended for use in the electromagnetic environment specifi ed below. The customer or user of the ATMOS C 451 should ensure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance...
  • Page 26 12.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS C 451 is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS C 451 should ensure that it is used in such an environment. IEC 60601- Electromagnetic Environment Immunity Test...
  • Page 27 ATMOS C 451 The ATMOS C 451 is intended for use in electromagnetic environment in which ratiated disturbances are con- trolled. The customer or user of the ATMOS C 451 can help prevent electromagnetic interference by maintai- ning a minimum distance between portable and mobile RF Communications equipment and the ATMOS C 451 as recommended below, according to the maximum output power of the communications equipment.
  • Page 28 Declaration of Conformity...
  • Page 29: General Standard Terms And Conditions

    ATMOS MedizinTechnik GmbH & Co. KG in accepting such goods or payment arrears. Acts of God or stoppages upon in writing by us.