Operation; Explanation Of The Display - Atmos C 051 Thorax Operating Instructions Manual

Table of Contents

Advertisement

4.0 Operation

4.1

Explanation of the display

Display of the
actual fistula
ATTENTION:
Actual vacuum
the current value. In the
Display of
case of a discontinuous
fistula, the value can
the actual
decline temporarily to zero
vacuum
although the fistula still
value.
exists.
Target vacuum
Display of the
preadjusted target
vacuum to which the
pump adjusts.
20
This is
Start / Stop
therapy
Key lock activated
Flow displayed as bubbles
Each additional coloured
bubble represents an
additional flow.
None: 0 - < 50 ml/min
Green: 50 - < 100 ml/min
Yellow: 100 - < 630 ml/min
Orange: 630 ml - < 2,01 l/min
Red: >2,01 l/min to maximum
Up to 1,00 l/min the flow is
displayed in ml/min.
Key lock manual
Day/Night Mode
The ATMOS® C 051 Thorax has a
day/night mode, the device adjusts
automatically to the environmental
lighting.
Under low ambient light conditions display
has dark background illumination.
System information
Therapy progress
Changeover to the
graphic diagram
User settings
Setting button
Increases the
target vacuum
Setting button
Decreases the
target vacuum