Atmos A 161 Operating Instructions Manual
Hide thumbs Also See for A 161:

Advertisement

English
ATMOS A 161 /
A 261 / C 161 /
C 261 Aspirator
GA1GB.310110.0
0124
2020-07 Index: 38

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A 161 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Atmos A 161

  • Page 1 English ATMOS A 161 / A 261 / C 161 / C 261 Aspirator GA1GB.310110.0 0124 2020-07 Index: 38...
  • Page 2: Table Of Contents

    Maintenance and servicing .......... 17 Suction procedure ............8 Reprocessing ............... 17 Hose rinsing ..............8 Sending in the device ........... 17 Operation ATMOS A / C 161 / 261 Aspirator / DDS ..9 10.0 Trouble-shooting ............18 5.1 DDS canister and bacterial filter ........9 Insertion of the DDS canister .........
  • Page 3: Introduction

    These operating instructions must always be kept available near the device. Care and period tests in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS A / C 161 / 261 Aspirator and are therefore a must besides regular cleaning.
  • Page 4: Function

    Introduction Function The ATMOS A / C 161 / 261 Aspirator is a very handy small suction unit. It is driven by an electromotive, maintenance-free swing piston pump. During operation, the pump generates a vacuum within the hose system and the collection canister, thus sucking off secretions or fluids (e.g. by means of a suction catheter).
  • Page 5: Extents Of Supply

    Scope of supply ● Prior to dispatch, this ATMOS device was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note). In addition to the basic device, the scope of delivery comprises the following parts:...
  • Page 6: Safety Advice

    For your Safety ● The safety standard of the ATMOS A / C 161 ● Prior to starting up the ATMOS A / C 161 / 261 / 261 Aspirator corresponds with recognized Aspirator, read these operating instructions carefully. medical technical regulations and the direc- tions of the law relating to medical products.
  • Page 7: Setting Up And Starting Up

    The unit may not be operated if it has not acclimatised as this might damage its diaphragms. ● ATMOS A / C 161 / 261 Aspirator / DDS: Always hold ready at least one ATMOS DDS bacterial filter, as the device may not be operated without one!
  • Page 8: General Operation Atmos A / C 161 / 261 Aspirator

    General operation Prior to these notes please read the foregoing chapter of your respective version of the ATMOS A / C 161 / 261 Aspirator! ● Attention: Suction procedures in the ● Check the vacuum readout regularly! respiratory tract may only be ● If secretion has been soaked up into the...
  • Page 9: Operation Atmos A / C 161 / 261 Aspirator / Dds

    Operation DDS ● Do not operate the device without a ● Exchange the bacterial filter when chang- bacterial filter. We recommend you ing the patient. ATMOS recommends: always to store at least one spare Replace the bacterial filter after 14 days, bacterial filter. even if there is no patient change. ● Always wear disposable gloves when ● Vacuum connection Direct-Docking-System...
  • Page 10: Operation Receptal

    Operation Receptal ® Holder for Receptal secretion ® canister Plug the screw thread of the holder from the top into the borehole at the bottom side of the device. Screw it together with the provided screw nut. Fig. 11. Assembling the Receptal ®...
  • Page 11: Operation Medi-Vac / Serres

    Operation Medi-Vac and Serres ® ® Holder for Medi-Vac secretion ® Medi-Vac ® canister Plug the screw thread of the holder from the top into the borehole at the bottom side of the device. Screw it together with the provided screw nut. Assembling the Medi-Vac ®...
  • Page 12: Cleaning/Disinfection

    8.1.3 Fingertip specialist, who is authorised by ATMOS, in compliance with Medical Devices Operator Ordinance, Medical Devices Act and ● Fingertip is not included in delivery, please order BV-Med rules is required (please see chapter 9.2 reprocessing).
  • Page 13: Device Surface

    ● Prior to use the device in another patient the Generally a reduced suction capacity is a sign of possible oversuction. The ATMOS C 161 / C 261 Aspirator have a condensate collector at complete device surface must always be cleaned the bottom of the device. For visual inspection please remove the with a damp (not wet) cloth and disinfected with a cap.
  • Page 14: Recommended Instrument Disinfections

    Non-ionic tensides (Application concentrate) Solvents, complexing agents Recommended disinfectants for surfaces Disinfectant Contents (in 100 g) Manufacturer ATMOS Green & Clean SK Alkyldimethylbenzylammoniumchloride <1 g Metasys, Rum (Austria) (application solution) Dialkyldimehtylammoniumchloride <1 g Alkyldimethylethylbenzylammoniumchloride <1 g...
  • Page 15: Cleaning And Servicing Plan

    Cleaning and servicing plan for ATMOS A 161 / A 261 / C 161 / C 261 Aspirator start date: name of the item: serial number: cleaning/exchange performed cleaning cleaning cleaning exchange of exchange of exchange secretion canister of the...
  • Page 16 Always fied ATMOS partner or an ATMOS authorized, specialized dealer. use the original ATMOS bacterial filter This cleaning and servicing plan as well as the relevant notes result ...
  • Page 17: Maintenance And Servicing

    Generally a reduced suction capacity is a sign of possible wards disinfect all secretion canister parts and hose oversuction. parts. The device´s surface also has to be disinfected. The ATMOS C 161 / C 261 Aspirator have a condensate collector at the bottom of the device (fig ). For visual ● ATMOS cannot guarantee perfect functioning neither inspection please remove the cap.
  • Page 18: Trouble-Shooting

    10.0 Trouble-shooting Prior to dispatch, the ATMOS A / C 161 / 261 Aspirator was subjected to an extensive functional test. If, nevertheless, a failure should appear, you may possibly clear it yourself if you follow these notes Problem Possible causes Remedy ● Unit does not start...
  • Page 19: Accessories, Consumables

    000.0836.0 Hose connector for hoses Ø 6 mm Grad. secretion canister 1 l 313.0007.0 Hose rewind for suction hoses for ATMOS A- and C- class Power cable, 3-wire Schuko connector, 2 m 008.0866.0 Carrying bag, black 313.0011.0 Device support for ATMOS A- and C- class 313.0012.0...
  • Page 20: Spare Parts

    11.0 Accessories, consumables and spare parts 11.3 Spare parts Hider ( Valid for ATMOS C 161 / C 261 Aspirator) 999.1996.0 Self-adhesive elastic buffer 5J-5017 000.0018.0 Plug GPN 300 000.0824.0 Ball for overflow protection 000.0839.0...
  • Page 21: Technical Specifications

    -80 kPa* ± 2 kPa -80 kPa* ± 2 kPa Vacuum readout -1...0 bar (± 25 mbar) (mmHg; kPa)* (ATMOS C 161 / C 261 Aspirator) Additional air regulation mechanical regulating valve (ATMOS C 161 / C 261 Aspirator) Additional air regulation...
  • Page 22: Checking / Reprocessing / Disposal

    Checking / Reprocessing / Disposal 13.1 Checking ATMOS suction devices The ATMOS suction devices are maintenance-free in the case they are used according to the operating instructions. Please comply with the country-specific guidelines regarding regular testing especially for the electrical safety.
  • Page 23: Notes On Emc

    14.1 Guidelines and Manufacturer´s Declaration - Emissions The ATMOS A / C 161 / 261 Aspirator is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS A / C 161 / 261 Aspirator should ensure that it is used in such an environment. Emissions Test Compliance...
  • Page 24 14.2 Guidelines and Manufacturer´s Declaration - Immunity for ATMOS A / C 161 / 261 Aspirator The ATMOS A / C 161 / 261 Aspirator is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS A / C 161 / 261 Aspirator should ensure that it is used in such an environment.. IEC 60601- Electromagnetic Environ-...
  • Page 25 To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured field strength at the location where the ATMOS A / C 161 / 261 Aspirator is used exceeds the above compliance level, the ATMOS A / C 161 / 261 Aspirator is to be observed to verify the intended use.
  • Page 26 The customer or user of the ATMOS C 161 Aspirator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS A / C 161 / 261 Aspirator as recommended below, according to the maximum output power of the communications equipment.
  • Page 27 For your notes...
  • Page 28 ATMOS MedizinTechnik GmbH & Co. KG Ludwig-Kegel-Str. 16 79853 Lenzkirch / Germany Phone: +49 7653 689-0 atmos@atmosmed.de www.atmosmed.com...

This manual is also suitable for:

A 261C 161C 261

Table of Contents