13.0 Notes on EMC (Electromagnetic compatibility)

Medical electrical equipment is subject to special precautions with regard to EMC and must
be installed acc. to following EMC notes.
• Portable and mobile HF communication facilities can influence medical electrical equipment.
• The use of other accessories, other power supplies and cables than stated may lead to an
increased emission or a reduced interference immunity of the equipment or system.
13.1

Guidelines and Manufacturer's Declaration - Emissions

The ATMOS® C 051 Thorax is intended for use in the electromagnetic environment specified
below. The customer or user of the ATMOS® C 051 Thorax should ensure that it is used in such
an environment.
Emissions Test
RF Emissions acc.to CISPR 11
RF Emissions acc.to CISPR 11
Harmonic emissions according
to IEC 61000-3-2
Voltage fluctuations/flicker
according to IEC 61000-3-3
52
Electromagnetic Environment -
Compliance
Guidance
Group 1
The ATMOS® C 051 Thorax uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
Class B
The ATMOS® C 051 Thorax is suitable
for use in all establishments, including
Class A
domestic, and those directly connected
to the public low-voltage power supply
Corresponds
network that supplies buildings used for
domestic purposes.