4.0
Operation
Fig 33
Fig 34
Fig 35
20
T R S
4.14.1
Monitoring the progression of therapy
via micro SD card
Via an integrated SD card, REF 312.2021.0, the stored
data can be output in Excel format and assigned to the
patient file. This makes it possible to trace the
progression of treatment at any later time as well.
!
Please only remove the SD Card when the device
is switched off.
At the beginning of each data logging session, the time and
date are recorded on the file. Subsequently, the flow and vacu-
um values are saved to the SD Card every 2 seconds. There-
fore the time for each of the recorded data samples can be
calculated by cumulatively adding 2 seconds to the start time.
4.15

Connection to the nurse call system

The cable may only be made by an electronic
specalist.
Prior to use the device, the nurse call system must be
!
checked (service menu item).
The ATMOS S 201 Thorax can be connected to the
nurse call system via the nurse call interface (NCI).
A stereo phone jack 3.5 mm is required. In the case of
an alarm the alarm signalling is transferred to the nurse
call system. During each alarm the contact at the phone
jack which is connected to the NCI jack T-S is closed
for approx. 1 s. The contact R-S is opened during this
time.
Check nurse call system (NCS):
Turn off device in off by pressing and holding ...
After device turned off press and
hold
and
until the ATMOS Logo appears.
The device enters into the service mode
Now check the menu item "NCS". It should say "yes".
If "NCS" status is "no", press
or
to change between "yes" and "no".
After this procedure just reboot device
simultaneously
to activate and press

Advertisement

Table of Contents
loading