Atmos C 051 Thorax Operating Instructions Manual page 33

Table of Contents

Advertisement

Display
WARNUNG
Batterie defekt
Das Gerät kann nicht betrieben werden!
Batterie durch Service tauschen lassen!
Cause of error
If the device is in a tilted
position the warning „Device in
critical tilt" appears.
If the therapy has not been
started after initial operation
the warning "Inactive therapy"
appears.
Defective battery. The device
can no longer be used.
Perform an inspection according to the manufacturer's specifications
every 12 months. This will be displayed on the device.
Low battery capacity appears in the display.
Battery must be replaced by the service.
If target vacuum is adjusted over -50 mbar the notice appears ,,High
target vacuum is set".
Troubleshooting
Place the device in an upright
position. The warning signal
is automatically deactivated.
Start therapy by pressing the
button.
Contact the ATMOS service!
33